BSD-2000 Deep Phased Array Commercial Product Comparative Table |
|
BSD-2000B |
BSD-2000-3D |
BSD-2000-3D/MR |
RF Generator |
Tetra |
Dodek |
Dodek |
# of RF Channels |
4 |
12 |
12 |
Directional Heat Zone Steering |
2 axis (X & Y) |
3 axis (X, Y & Z) |
3 axis (X, Y & Z) |
Maximum Power |
1300 |
1800 |
1800 |
Max. Channel Power |
325 |
450 (4) or 135 (12) |
135 |
Frequency (MHz) |
75 to 140MHz |
75 to 140MHz |
100MHz |
# of Antennas |
8 |
8 and 24 |
24 |
Forward Pwr Cont. |
Yes |
Yes |
Yes |
Phase Control |
Yes |
Yes |
Yes |
MRI Compatible |
No |
No |
Yes |
Multiple size applicators |
Yes |
Yes |
Yes |
Computer Console |
included |
included |
included |
Sigma 30 (30cm) applicator |
Yes |
Yes |
No |
Sigma 60 (60cm) applicator |
Yes |
No |
No |
Sigma Ellipse applicator |
Yes |
No |
No |
Sigma Eye applicator |
No |
Yes |
No |
Sigma 30MR applicator |
No |
No |
Yes |
Sigma Eye MR applicator |
No |
No |
Yes |
# TemperatureProbes |
8 |
8 |
8 |
# Sensors per probe |
1 |
1 |
1 |
# of Sensor Locations Along Catheter |
31 |
31 |
31 |
Thermal Mapping |
Yes |
Yes |
Yes |
Built-in Thermal Calibration |
Yes |
Yes |
Yes |
CE Mark |
Yes |
Yes |
Yes |
FDA Clearance in U.S. |
Yes HDE |
in process |
in process |
Forward Pwr. Detection |
Yes |
Yes |
Yes |
Reflected Pwr. Detection |
Yes |
Yes |
Yes |
Water Heating/Cooling System |
Built-in |
Built-in |
External |
Water Bolus Interface |
Yes |
Yes |
Yes |
Variable Focal Diameter |
Yes |
Yes |
Yes |
Non-invasive thermometry |
No |
No |
Yes |
Published Clinical Trials |
Yes |
Yes |
Yes |
# of Water Control Systems Needed |
1 |
1 |
1 |
ESHO Guideline Compliant |
Yes |
Yes |
Yes |
Automatic Bolus Air Purge |
Yes |
Yes |
Yes |
UV Water Treatment |
Yes |
Yes |
Yes |
Integral Deionizer of Water |
Yes |
Yes |
No |
Currently Available For Sale |
Yes |
Yes |
Yes |
Pretreatment Planning |
Yes |
Yes |
Yes |
Shield Room Required |
Yes |
Yes |
Uses Existing MRI Shielding |